A9800

Country:
USA
Standard:
FDA,CE,EN 60601-1, EN 60601-1-2, ISO 80601-2-13
Category:

Description

15″TFT LCD with touch screen along with a navigator wheel provides a simple intuitive interface that enhances user control.

Automatic Compliance compensation along with Fresh Gas Flow compensation help clinician to deliver accurate and precise ventilation therapy.

Full waveform display including integrated Spirometry provides loops when desired for improved clinical data analysis.

System provides a minimum of 25% O2 concentration at all times utilizing a pneumatic Oxygen Ratio Controller. This enhances patient safety over systems that utilize electronic or software controlled ORC functions.

Large stainless worktable provides extra convenience to the user, with the flipup table design saving space efficiently.

Impressive array of standard features improves the system usability: auxiliary oxygen flow meter with Air/O2 Blender and auxiliary AC power outlets.

Air/O2 Blender for Aux Gas outlet is conveniently located on front panel to allow user to mix Aux Gas from 100% O2 to 21% O2 easily and accurately.

Breathing system is heated to reduce condensation and is integrated within the workstation.

Advanced breathing system is fully autoclaveable.

Absorber by-pass function makes changing absorbent during a case easier.

Absorber is compatible with Standard Prepaks or loose-fill CO2 Absorbent.

Auto/Manual Switch is located on the breathing system for simple control of Ventilator On/Off.

ACGO (Auxiliary Common Gas Outlet) switch is a standard feature to allow for use with non-rebreathing adapters.

GCX-type rails support easy mounting of other devices to the workstation.

6 Traditional Gas Supply Pressure Gauges allow for easy status monitoring of hospital wall gas supplies and gas cylinders/tanks.

3 locking drawers for storage.

Standard Selectatec™-compatible mounts hold two vaporizers.


Ventilation Modes
Volume Control
Pressure Control
SIMV with Pressure Support (SIMV + PS) – with Volume or
Pressure-type mandatory breaths
Pressure Support (with Apnea backup)
PCV-VG (Pressure Control Ventilation with Volume Guarantee)
Manual ventilation

Parameters and Ranges
Pressure target: 5 – 70 cmH2O
Pressure support (delta P): 3 – 50 cmH2O
Tidal Volume: 20 – 1500 mL
Breathing Freq
2-100 bpm
2-60 bpm in PS
2-100 bpm in SIMV-VC and SIMV-PC
4-100 bpm in other modes
TINSP: 0.2-5.0 s
PEEP: OFF, 3 – 30 cmH2O
FreqMIN: 2 – 60 bpm
T pause: OFF, 5%-60%
Trigger: 1 – 15 L/min
I:E Ratio: 4:1-1:8
TSLOPE: 0 – 2 s
Vaporizers: Sevoflurane, Desflurane, Isoflurane, Halothane
Enflurane

Monitoring
Continous monitoring of inspiratory O2 concentration, breathing
frequency, airway pressure (Paw, Peak, Pmean, PEEP), minute
volume and tidal volume
The measured parameters are displayed as large, easy to read
digital values. Airway pressure, flow and CO2 (optional) are shown
in graphical waveforms

Alarms
Apnea
Apnea CO2
Adjustable alarm limits for Inspiratory O2concentration (FiO2)
Adjustable alarm limits for Minute Volume (MV)
Adjustable alarm limits for Airway pressure (PAW)
Adjustable alarm limits for EtCO2and agents
Continuous Pressure
O2Supply Fail
Negative Pressure
High Breath Rate (PS)
High PEEP
Mixed Agents
Low Battery
AC Power Fail
Technical alarms

Operation Conditions
Operating voltage: AC100-240V, 50Hz/60Hz
Temperature: 10 – 40°C(operation); -20 – 60°C(storage &transport)
Relative humidity: ≤90%, non-condensing (operation)
≤90%, non-condensing(storage &transport)
Weight (without vaporizer & cylinders): approx. 100Kg
Dimensions (H x W x D): approx. 1400mm x 900mm x 760mm

Standards
EN 60601-1, EN 60601-1-2, ISO 80601-2-13